UDAF HOOD

UDAF HOOD

Vertical flow

The vertical LAF laminar-flow hood is designed for use when handling non-active excipients in powder form, ensuring containment of the particulate in a confined space, while also safeguarding the operator. The entire system complies with GMP regulations and guarantees a controlled environment, even during the most critical stages of the process.

Central column
Upper column
Control system

Central COLUMN

The central column is manufactured from sandwich panel with AISI 304 stainless steel sheet and Scotch Brite finish, with low surface roughness (Ra < 1.6 μm) and aluminum honeycomb interior.

All components are coplanar, with rounded edges and easy-to-sanitize surfaces. The rear casing houses the suction grids with G4 filters, the canisters with F9 filters, the HMI panel, complete with emergency stop buttons, pressure switches for monitoring the filters, and the electrical panel.

The instrumentation is accessed from the rear of the machine via the technical area, however the hood is also available in a compact version with complete frontal access, upon request. All joints are sealed with silicone to ensure airtightness and hygiene.

Upper Column

The upper casing is manufactured from AISI 304 stainless steel sheet sandwich wall panel and Scotch Brite exterior finish, with low surface roughness (< Ra 1.6 µm) and aluminum honeycomb filling.

All components are coplanar with rounded edges and easy to clean surfaces.

The upper casing houses the following elements:

  • The delivery fan that feeds the absolute filters and the expulsion filters
  • The H14 absolute delivery filters (box type).
  • The H14 absolute filters for the air expelled into the room (box type).
  • The LED lamps for indoor lighting

The delivery filters, expulsion filters and LED lamps may be replaced from below. All joints are sealed with silicone to ensure airtightness.

The perimeter compartmentalization system, consisting of strips of flexible, transparent PVC, which, in the standard version, permits access to the working area via the front only, is mounted on the upper casing.

Control system

The system devices, equipment and components are controlled by PLC, which enables the operator to manage and monitor the hood system via the touch-screen operator interface.

The main functions of the control system are:

  • Fan management
  • Filter and fan alarm indications
  • Graphical interface

The system features 2 operating modes:

  1. Local mode (standard): the machine can be started and stopped in manual mode by the operator.
  2. Remote mode (optional): the machine starts automatically when the department HVAC system is running.

The ventilation system operates in partial recirculation mode, drawing air in from the working area, conveying it through a series of progressive filters and returning it, purified, so as to maintain the integrity of unidirectional flow and a constant internal vacuum.

The air passes through a triple filtration stage consisting of G4 pre-filters, F9 canister filters with BIBO system, and HEPA H14 absolute filters, ensuring effective removal of particles and contaminants before it is re-introduced into the environment.

MODELS
Mod. DCFH-01
Delivery line filtration 4 x HEPA H14 filters, dim. 610x610x150
Delivery line flow rate 0.45 m/s ± 20% at 150 mm from the filter
Airflow management Incoming flow from the room to the hood
Return line filtration 2 x G4 pre-filters, dim. 592x592x48
1 x F9 pre-filter (in BIBO canister) dim. 592x592x287
Expulsion line filtration No. 2 x HEPA H14 filters, dim. 610x305x150
Delivery line ventilation Centrifugal electric fan with inverter
Lighting 4 x LED lamps in recesses on the UDAF
Compartmentalization Flexible, transparent strips of PVC, complete with
AISI 304 stainless steel frame
Instruments Differential pressure transmitters for monitoring filter clogging
Differential pressure transmitters for PID fan control
Environmental classification under udaf Grade A and ISO 5 in standby
Grade C and ISO 8 when operating
Installation Floor and ceiling mounting
Mod. DCFH-02
Delivery line filtration 4 x HEPA H14 filters, dim. 610x610x150
Delivery line flow rate 0.45 m/s ± 20% at 150 mm from the filter
Airflow management Incoming flow from the room to the hood
Return line filtration 2 x G4 pre-filters, dim. 592x592x48
1 x F9 pre-filter (in BIBO canister) dim. 592x592x287
Expulsion line filtration No. 2 x HEPA H14 filters, dim. 610x305x150
Delivery line ventilation Centrifugal electric fan with inverter
Lighting 4 x LED lamps in recesses on the UDAF
Compartmentalization Flexible, transparent strips of PVC, complete with
AISI 304 stainless steel frame
Instruments Differential pressure transmitters for monitoring filter clogging
Differential pressure transmitters for PID fan control
Environmental classification under udaf Grade A and ISO 5 in standby
Grade C and ISO 8 when operating
Installation Floor and ceiling mounting

Each structural element is designed to guarantee hygiene and ease of cleaning: the central column and upper casing are made of AISI 304 stainless steel with Scotch Brite finish and low roughness sandwich panels with coplanar surfaces and rounded corners.

The working area is delimited by transparent PVC strips, supported by a stainless steel frame, which permits access via the front of the machine in the standard version, while guaranteeing effective separation between the indoor and outdoor environments. The entire system is supervised by a PLC, with a touch screen operator interface, which enables automatic fan control, alarm indications, and management of local or remote operation in synchrony with the HVAC system.

TECHNICAL DATA
DESCRIPTION UNIT DCFH-01 DCFH-02
Part No. MCCAPPA0001 MCCAPPA0005
External dimensions WxDxH1 mm 1470x2330x3630 1750x2440x3720
Dimensions of the working area WxDxH mm 1410x1350x2225 1750x1450x2230
Weight kg 530 660
UDAF zone air flow rate m3/h 2800 4300
Expulsion zone air flow rate m3/h 1400 1050
Noise level dB(A) < 70 < 70
Lighting lux 750 750
Electrical power supply 400 V 3P+N+E @50 Hz
Electrical power kW 4.4 4.7
Power consumption 2 A 3.7 4
¹ Before shipping, the machine is divided into smaller sub-assemblies to make it easy to lift, handle and move
² Clean filters and lifting active